Case 04 · Right side
Right fronto-temporo-parietal cranioplasty
Extensive right lateral vault defect, large perforated implant.
Loading the model · 0 %
View
Layers
Section plane
Measurement
Turn on measurement, then click two points on the bone, the implant or a screw. Distances are in millimetres, on the manufacturing meshes.
Implant section in blue, bone section in ochre. Models simplified for the web: measurements are indicative; the reference dimensions are those of the design file.
Reading the case
What the model shows
A large one-piece titanium implant, 104 mm long and 100 mm high, rebuilding the right side of the vault. Its outline follows the bony margin of the defect. A regular grid of 59 perforations lightens the part and encourages drainage and tissue adhesion; eleven small holes around the edge are intended for fixation. The identification engraving sits at the centre of the outer face. Screws are not modelled in this case.
Read the case reviewWhat to look at in the viewer
- 01Transparent bone: the implant outline follows the margin of the defect all the way round.
- 02Axial or coronal section plane: plate thickness and continuity of the curvature with the healthy vault.
- 03Implant moved away: shape and extent of the fronto-temporo-parietal defect.
- 04Measurement: diameter of a perforation, distance between two fixation holes.
Measured values
- Region
- Cranial vault
- Overall size
- 104 × 100 × 34 mm
- Thickness
- ≈ 2.2 mm
- Volume
- 17.6 cm³
- Mass (Ti-6Al-4V)
- ≈ 78 g
- 59 perforations Ø ≈ 3.8 mm
- 11 fixation holes Ø ≈ 1.8 mm
- Identification engraving
Measurements taken on the manufacturing meshes. Thickness: median distance between opposite faces; mass: volume × 4.43 g/cm³.
Augmented reality · 1:1 scale
The implant at full size, on your table
Place the implant in the room at its real dimensions (104 × 100 × 34 mm) and walk around it with your phone.
A similar case to plan?
Send the imaging: the feasibility study comes before any design.